Configurable with turbomolecular pump speeds ranging from 47 to 400 ls-1. ... And a choice of oil sealed or dry backing pumps ranging from 1 to 20 m³h-1.
They all feature an integrated TIC turbo and instrument controller offering full control of the package via a simple intuitive interface. The nEXT turbomolecular pumping stations are supplied fully assembled and ready to run straight out of the box and include common accessories such as mist filters and mains cables as appropriate to the chosen pumps. As fully featured high end stations they include RS232 serial communications and Windows® software for monitoring and control.
Features and Benefits
Range of turbomolecular pump options: choice of turbomolecular pump with speeds ranging from 47 to 400 l/s and inlet flanges from DN40 to DN160.
Fully controllable: TIC turbo and instrument controller offers full control of pumps and up to 3 Active gauges as well as offering full serial remote communications.
Vent valve option: optional turbomolecular pump vent valve can be ordered as part of cart assembly.
User serviceable: all nEXT turbomolecular pumps and backing pumps are fully user serviceable.
Robust metal frame: all metal frame with locking castors for a robust but easily mobile system. Bench mounting kit included for safe bench top operation.
Choice of backing pumps: oil sealed or dry backing pumps with capacities ranging from 1 to 20 m³/h
Low vibration: backing pump mounted on anti-vibration mounts for low levels of transmitted vibration.
The dimensions length, width and height try to describe the outer shape of this particular product. We always recommend checking the detailed dimensions given in 2D drawings in the instruction manual or the 3D shapes given in 3D models.
Certification Motor
CSA (C22.2 No.61010-1-12)
Performance Data
Pressure Ultimate
6.000 × 10⁻⁶
Nominal rotational speed
1000.00
Power at Ultimate
20.00
Pumping Speed Air Nitrogen (@50Hz or max @default setting)
0.300
Pumping Speed Helium
0.340
Pumping Speed Hydrogen
0.280
Environmental Conditions
Temperature Ambient Operation Max
313.15
Temperature Ambient Operation Min
285.15
Temperature Ambient Storage Min
243.15
Temperature Ambient Storage Max
343.15
Magnetic Field Resistance Max (covering worst direction - come and consult us for more details)